Tau ELISA Kit Wako Technical Information and Data
Performance


Standard curve range | 4.10 pg/mL - 1,000 pg/mL |
Reactive Tau | Total Tau |
Sample | Human cerebrospinal fluid (CSF) |
Sample volume | 10 µL~※1 |
Assay time | 3 hours |
Detection method | Luminescent system※2 |
※1: It is recommended that CSF sample volume is 50 μL, considering the accuracy of dilutions.
※2: A luminescence plate reader is required for the assay.
Principles
This kit is sandwich ELISA measuring phosphorylated Tau T181 by using 2 monoclonal antibodies. One recognizes phosphorylated Tau T181, and another reacts middle region of Tau.


Tau isoforms


Phosphorylation sites of Tau


Application data
Spike-and-recovery and dilution linearity
Spike-and-recovery
Human CSF 1 | Not spiked | + 20 pg/mL Tau | + 60 pg/mL Tau | + 150 pg/mL Tau | |||
---|---|---|---|---|---|---|---|
Measured | Measured | Recovery | Measured | Recovery | Measured | Recovery | |
4.09 pg/mL | 24.8 pg/mL | 104% | 58.1 pg/mL | 90% | 175 pg/mL | 114% | |
Human CSF 2 | Not spiked | + 40 pg/mL Tau | + 135 pg/mL Tau | + 380 pg/mL Tau | |||
Measured | Measured | Recovery | Measured | Recovery | Measured | Recovery | |
24.8 pg/mL | 70.5 pg/mL | 114% | 160 pg/mL | 100% | 390 pg/mL | 96.1% |
Dilution linearity


These assays demonstrated that the kit is highly quantitative.
Measurements of human CSF
Cerebrospinal fluids(CSF) from patients were measured with our ELISA kit and competitor’s kit.
The samples are from Alzheimer disease (AD) , mild cognitive impairment(MCI) and non-dementia patients.
The differences between non-dementia, MCI, and AD groups were more significant when measured with our kit.




Correlation with Mini Mental State Examination (MMSE) scores


※MMSE score
≤ 23: suspected of dementia
24-27: suspected of mild cognitive impairment
28-30: normal


The correlation between MMSE scores and the concentration of Tau was examined. Samples were collected from CSF and measured using our ELISA kit.
There was a correlation between the concentration of Tau and MMSE score. The correlation was stronger when Tau concentration was measured using our product.