LabAssay LDL-Cholesterol Technical Information and Data

Kit Performance

Sample Serum, Plasma
Target animals Mouse, Rat
Calibration curve range 9.38-300 mg/dL
Sample volume 5 μL
Measurement duration Approx. 20 min
Wavelength Primary wavelength 600 nm
Reference wavelength 700 nm

Example of Calibration Curve


Principle

1) The first reaction(elimination of non-LDL cholesterol)

The block polymer containing hydrophilic and hydrophobic moiety in Pretreatment selectively binds to LDL in the specimens, and protects it from enzyme (CHE, CO) reactions. CHE and CO react with non-LDL lipoprotein [chylomicron (CM), very low density lipoprotein (VLDL), high density lipoprotein (HDL)]. Hydrogen peroxide produced by the enzyme reactions with non-LDL cholesterol is decomposed to water by catalase in Pretreatment reagent.

2) The second reaction(color reaction of LDL-cholesterol)

When Reacting Solution is added, the cholesterol and its derivatives in LDL produce hydrogen peroxide by CHE and CO. Hydrogen peroxide produced by the enzyme reactions with LDL-cholesterol yields a blue color pigment upon oxidative condensation with N-(3-sulfopropyl)-3-methoxy-5-methylaniline (HMMPS) and 4-aminoantipyrine in the presence of peroxidase (POD). The amount of LDL-cholesterol contained in the sample is determined by measuring the absorbance of the blue color.


Data

Precision of assay (within assay variation)

Reproducibility was confirmed by performing quintuplicate assays of serum samples of two concentrations each from humans, mice, and rats with this product.

n\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL)
1 160 46.0 141 48.3 119 33.3
2 156 43.8 138 47.9 114 34.1
3 158 44.7 138 49.5 119 33.3
4 166 44.5 139 48.5 112 34.9
5 159 45.7 136 50.6 113 33.3
mean 160 44.9 138 49.0 115 33.8
SD 3.8 0.90 1.8 1.1 3.4 0.72
CV (%) 2.4 2.0 1.3 2.2 2.9 2.1

 

Result: The CV (%) of human serum was 2.0-2.4%, mouse serum was 1.3-2.2%, and rat serum was 2.1-2.9%, indicating good reproducibility.


Reproducibility (between assay variation)

Triplicate measurements were performed on human, mouse, and rat serum samples at three concentrations each for four days with this product.

Day\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL)
1 125 88.1 41.8 119 91.4 48.4 130 78.7 45.2
2 124 87.0 41.1 119 88.1 45.0 129 76.1 42.9
3 130 91.5 41.7 118 88.0 48.2 126 74.2 49.7
4 125 89.5 39.8 126 95.1 50.4 131 74.5 49.4
mean 126 89.0 41.1 121 90.7 48.0 129.0 75.9 46.8
SD 2.7 1.9 0.92 3.7 3.4 2.2 2.2 2.1 3.3
CV (%) 2.1 2.2 2.2 3.1 3.7 4.7 1.7 2.7 7.1

 

Result: The CV (%) of human serum was 2.1-2.2%, mouse serum was 3.1-4.7%, and rat serum was 1.7-7.1%, indicating good reproducibility.


Dilution linearity test

Using the standard diluent supplied with the product, human serum/plasma (EDTA), mouse serum/plasma (EDTA), and rat serum/plasma (EDTA) at two different concentrations each, were subjected to a 2-fold serial dilution with sample buffer and measured to confirm linearity (measured in duplicates).

Result: Good linearity was confirmed for all samples within the calibration curve range.


Spiked recovery test

Standard solutions at three concentrations were added to human serum/EDTA plasma, mouse serum/EDTA plasma and rat serum/EDTA plasma. Next, spike-recovery tests were conducted, with each measurement performed in duplicate.

Human

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 75.2 - -
7.12 82.3 7.10 100
28.5 103 27.8 97.5
35.6 112 36.8 103
Plasma
(EDTA)
- 84.5 - -
10.6 93.8 9.30 87.7
21.2 107 22.5 106
26.5 112 27.5 104

Mouse

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 86.6 - -
15.9 102 15.4 96.9
23.8 112 25.4 107
39.7 126 39.4 99.2
Plasma
(EDTA)
- 97.7 - -
14.4 112 14.3 99.3
28.9 126 28.3 97.9
36.1 133 35.3 97.8

Rat

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 14.6 - -
13.7 28.7 14.1 103
41.0 53.0 38.4 93.7
68.4 77.1 62.5 91.4
Plasma
(EDTA)
- 16.2 - -
16.0 31.9 15.7 98.1
48.1 64.4 48.2 100
80.2 90.6 74.4 92.8

 

Result: Good recovery rates was confirmed for all samples.


Measurement examples

Human serum/plasma, mouse (ICR) serum/plasma samples, and rat (SD) serum/plasma were measured with this product.

Human

  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 189 187 188 1.4 0.75
2 88.8 93.2 91 3.1 3.4
3 85.7 85.7 85.7 0 0
4 99 93.9 96.5 3.6 3.7
5 95.8 94 94.9 1.3 1.3
Plasma
(EDTA)
1 122 121 122 0.71 0.58
2 88.3 88 88.2 0.21 0.24
3 85.9 84.4 85.2 1.1 1.2
4 93.9 97.2 95.6 2.3 2.4
5 95.1 95.4 95.3 0.21 0.22

Mouse

  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 32.6 33.4 33 0.57 1.7
2 33.6 31.3 32.5 1.6 5
3 33.4 32.4 32.9 0.71 2.1
4 34.5 33.4 34 0.78 2.3
5 36.1 34.1 35.1 1.4 4
Plasma
(EDTA)
1 21.2 20.8 21 0.28 1.3
2 16 17.7 16.9 1.2 7.1
3 25.5 21.2 23.4 3 13
4 19.1 20.3 19.7 0.85 4.3
5 18.9 19.6 19.3 0.49 2.6
 
Rat
  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 43.3 37.9 40.6 3.8 9.4
2 44.4 42.3 43.4 1.5 3.4
3 42.7 44 43.4 0.92 2.1
4 44.9 43.2 44.1 1.2 2.7
5 50 51.1 50.6 0.78 1.5
Plasma
(EDTA)
1 30.7 30.1 30.4 0.42 1.4
2 33.7 32.8 33.3 0.64 1.9
3 39.2 36.5 37.9 1.9 5
4 30.7 30.9 30.8 0.14 0.46
5 33.9 36.7 35.3 2 5.6
© 2025 FUJIFILM Irvine Scientific. All rights reserved.