LabAssay HDL-Cholesterol Technical Information and Data

Performance

Sample
Serum, Plasma
Target animals
Mouse, Rat
Calibration curve range
6.25-200 mg/dL
Sample volume
5 μL
Measurement duration
Approx. 20 min
Wavelength
Primary wavelength 600 nm
Reference wavelength 700 nm
 

Example of Calibration Curve


Principle

1) The first reaction (elimination of non-HDL cholesterol)

The block polymer containing hydrophilic and hydrophobic moiety in Pretreatment selectively binds to HDL in the specimens, and protects it from enzyme (CHE, CO) reactions. CHE and CO react with non-HDL lipoprotein [chylomicron (CM), very low density lipoprotein (VLDL), low density lipoprotein (LDL)]. Hydrogen peroxide produced by the enzyme reactions with non-HDL cholesterol is decomposed to water by catalase in Pretreatment reagent.


Data

Precision of assay (within assay variation)

Reproducibility was confirmed by performing quintuplicate assays of serum samples of two concentrations each from humans, mice, and rats with this product.

n\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL)
1 82.3 33.3 72.0 21.2 55.0 21.4
2 82.5 32.2 73.2 20.7 53.5 21.5
3 82.2 32.9 72.1 21.2 54.0 21.1
4 81.9 33.4 71.4 22.1 54.3 22.8
5 82.2 32.9 72.8 21.3 55.8 22.1
mean 82.2 32.9 72.3 21.3 54.5 21.8
SD 0.22 0.47 0.71 0.50 0.90 0.68
CV (%) 0.26 1.4 0.98 2.4 1.7 3.1

 

Result: The CV (%) of human serum was 0.26-1.4%, mouse serum was 0.98-2.4%, and rat serum was 1.7-3.1%, indicating good reproducibility.


Reproducibility (between assay variation)

Triplicate measurements were performed on human, mouse, and rat serum samples at three concentrations each for four days with this product.

Day\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL)
1 79.2 47.0 25.2 77.7 44.6 26.3 56.0 39.7 19.0
2 77.6 45.4 23.6 74.2 42.2 25.2 58.4 39.6 18.2
3 80.5 48.8 24.8 79.9 45.2 26.8 60.5 41.5 19.0
4 79.2 46.9 25.1 80.5 44.4 26.5 59.1 40.5 18.8
mean 79.1 47.0 24.7 78.1 44.1 26.2 58.5 40.3 18.8
SD 1.2 1.4 0.74 2.9 1.3 0.70 1.9 0.88 0.38
CV(%) 1.5 3.0 3.0 3.7 3.0 2.7 3.2 2.2 2.0

 

Result: The CV (%) of human serum was 1.5-3.0%, mouse serum was 2.7-3.7%, and rat serum was 2.0-3.2%, indicating good reproducibility.


Dilution linearity test

Using the standard diluent supplied with the product, human serum/plasma (EDTA), mouse serum/plasma (EDTA), and rat serum/plasma (EDTA) at two different concentrations each, were subjected to a 2-fold serial dilution with sample buffer and measured to confirm linearity (measured in duplicates).

Result: Good linearity was confirmed for all samples within the calibration curve range.


Spiked recovery test

Standard solutions at three concentrations were added to human serum/EDTA plasma, mouse serum/EDTA plasma and rat serum/EDTA plasma. Next, spike-recovery tests were conducted, with each measurement performed in duplicate.

Human

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 39.5 - -
11.3 50.3 10.8 95.6
22.5 61.8 22.3 99.1
28.2 67.1 27.6 97.9
Plasma
(EDTA)
- 36.0 - -
7.72 43.8 7.80 101
15.4 50.4 14.4 93.5
19.3 54.3 18.3 94.8

 

Mouse

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 11.7 - -
11.8 22.5 10.8 91.5
35.4 47.0 35.3 99.7
59.1 67.4 55.7 94.2
Plasma
(EDTA)
- 10.5 - -
9.34 19.2 8.70 93.1
28.0 36.6 26.1 93.2
46.7 54.9 44.4 95.1

 

Rat

  Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 15.6 - -
15.6 30.0 14.4 92.3
31.2 45.0 29.4 94.2
39.1 51.4 35.8 91.6
Plasma
(EDTA)
- 13.0 - -
13.2 25.3 12.3 93.2
19.9 32.8 19.8 99.5
33.1 44.0 31.0 93.7

 

Result: Good recovery rates was confirmed for all samples.


Measurement examples

Human serum/plasma, mouse (ICR) serum/plasma samples, and rat (SD) serum/plasma were measured with this product.

Human

  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 58.2 57.9 58.1 0.2 0.37
2 44.1 45.2 44.7 0.8 1.7
3 45.7 45.7 45.7 0.0 0.0
4 35.3 34.3 34.8 0.7 2.0
5 34.9 34.3 34.6 0.4 1.2
Plasma
(EDTA)
1 104 105 105 0.71 0.68
2 44.7 45.5 45.1 0.57 1.3
3 46.2 46.4 46.3 0.14 0.31
4 34.5 35.3 34.9 0.57 1.6
5 33.4 33.9 33.7 0.35 1.1

Mouse

  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 74.1 76.2 75.2 1.5 2.0
2 71.9 71.7 71.8 0.14 0.20
3 77.2 80.3 78.7 2.2 2.8
4 86.5 87.7 87.1 0.85 0.97
5 68.2 69.3 68.7 0.78 1.1
Plasma
(EDTA)
1 40.6 40.3 40.5 0.21 0.52
2 25.5 26.0 25.8 0.35 1.4
3 50.1 48.6 49.4 1.1 2.1
4 27.2 28.9 28.1 1.2 4.3
5 24.2 25.7 25.0 1.1 4.3
 
Rat
  No. Measured value (mg/dL) mean (mg/dL) SD CV (%)
Serum 1 22.5 20.5 21.5 1.4 6.6
2 31.8 30.7 31.3 0.78 2.5
3 32.7 33.2 33.0 0.35 1.1
4 22.4 22.5 22.5 0.071 0.31
5 25.5 25.8 25.7 0.21 0.83
Plasma
(EDTA)
1 13.2 13.3 13.3 0.071 0.53
2 13.3 13.1 13.2 0.14 1.1
3 13.1 11.4 12.3 1.2 9.8
4 13.6 13.6 13.6 0.0 0.0
5 15.4 16.4 15.9 0.71 4.4