Join Our Mailing List Today!

NOTICE: Online ordering will be disabled 1/27-2/3. Learn More.

Human β Amyloid (1-42) ELISA Kit Wako, High Sensitive Data

Performance

Standard curve

Standard curves were prepared using the standard solution included in the kit. The curves resembled a straight line between 0.1-20.0 pmol/L.


Spike-and-recovery test

Standard curves were prepared using the standard solution included in the kit. The curves resembled a straight line within the range between 0.1-20.0 pmol/L.

※2-hr reaction with the HRP-conjugated antibody


Dilution linearity test

Normal Human Plasma (A, C, E) were diluted 2-16 folds to determine the dilution linearity.

Specificity test

Sample Cross-reactivity (%)
Human Aβ (1-40) ≦ 0.1
Human Aβ (1-42) 100.0
Human Aβ (1-43) 13.5
Rat (Mouse) Aβ (1-40) ≦ 0.1
Rat (Mouse) Aβ (1-42) 0.5

 


Application Data

Measurements in human and mouse plasma

Blood samples collected in evacuated blood collection tubes with K2EDTA were centrifuged for 15 min. at 5,000 x g and 4°C. Plasma was isolated and kept in -80°C until use.

The plasma samples were diluted 4-fold using the standard diluent solution contained in the kit prior to the measurements.

(※Human β Amyloid(1-40) ELISA Kit Wako II (Catalog ID 298-64601) and Human/Rat β Amyloid(40) ELISA Kit Wako II (Catalog ID 294-64701) were used for the measurements of Aβ [1-40] and Aβ [x-40]).

 

(Data provided by: Drs. Iwatsubo and Hashimoto, Department of Neuropathology, Graduate School of Medicine, University of Tokyo)

© 2025 FUJIFILM Irvine Scientific. All rights reserved.